CP305-IO Touch-Control Panel 5"
Product description
The CP305 is a universal signaling and testing combination for medical and other areas. Messages and faults are displayed in accordance with the requirements of IEC60364-7-710 and NIN 7.10.
Required alarms are indicated visually and acoustically.
Important display functions:
- Normal operation display
- Insulation fault
- Overload
- Overtemperature
- Message from the changeover devices
- Messages from other Bender devices (EDS/RCMS)
- Device failures
- Test results from medical IT system
- Measured value display
Device variants:
CP305
The CP305 is used to display visual and audible alarms. ISOMETER® can be tested using the test function. The CP305 can also be used as a parallel display
with other CP305 or CP9xx.
In the event of an alarm, the programmed message texts are shown on the display.
CP305-IO
In addition to the CP305, the CP305-IO has 12 digital inputs that can be freely assigned
can be freely assigned. Messages from external devices such as battery systems or the status of medical gases can be reported in this way.
Any message texts can be assigned to the inputs.
The CP305-IO has 2 additional relay outputs.